Table 1:
Performance metric |
Parameter measured |
Mask airflow direction |
Measurement instrument |
Measurement method |
Particle size | Instrument airflow rate (liters/minute) |
---|---|---|---|---|---|---|
Cough aerosol collection efficiency | Aerosol mass escaping beyond mask into collection chamber | Outward | Andersen impactor | Fluorescence of aerosol on impactor stages | 0.1 to 7 μm | 28.3 |
Exhaled aerosol collection efficiency | Aerosol mass escaping beyond mask into collection chamber | Cyclical | Andersen impactor | Fluorescence of aerosol on impactor stages | 0.1 to 7 μm | 28.3 |
Filtration efficiency | Aerosol mass concentration upstream & downstream of mask | Inward | TSI 8130A | Photometer | 75 nm count median diameter (GSD 1.86) | 85 |
Airflow resistance | Pressure required for 85 lpm flow | Inward | TSI 8130A | Manometer | n/a | 85 |
Manikin fit factor | Aerosol particle count inside & outside mask | Cyclical | TSI PortaCount | Size classifier & condensation particle counter | 55 nm (N95 mode) | 0.35 |
Human fit factor | Aerosol particle count inside & outside mask | Cyclical | TSI PortaCount | Size classifier & condensation particle counter | 55 nm (N95 mode) 0.02 – 1.0 μm (Class 100 mode) | 0.35 |