Published in final edited form as: Noise Control Eng J. 2019 Jan 1;67(1):23–30. doi: 10.3397/1/37673

Table 2—

Measurement differences between the NIOSH SLM app and type 1 SLM by external microphone type.

Microphone Metric (dB) Laboratory test Field test
N Mean SD N Mean SD
i436 L Aeq 64 1.1 1.6 6 −0.8 1.4
L max 64 1.9 0.9 6 −1.5 2.4
L peak 64 0.3 1.2 6 1.1 1.5
iMM6 L Aeq 66 −0.5 0.9 14 −2.6 2.6
L max 66 0.6 1.7 14 −2.5 2.8
L peak 66 −0.5 5.0 14 0.0 2.1