Technical aspects of dynamic CZT-SPECT studies included in meta-analysis

Authors Index test Mean total effective dose (mSv) Rest dose (MBq) Stress dose (MBq) Stress injection protocol CZT cut-off value
Agostini et al.10 Rest/stress Tc MPI 8.73 3MBq/kg 9MBq/kg Regadenoson (400 μg) MFR < 2.1
Bouallègue et al.15 Rest/stress Tc MPI 8.4 220-280 645-730 Dipyridamole 0.75 mg/kg MFR < 2
Han et al.16 Stress Tl/rest Tc MPI 9.13 222–370 44–66 Adenosine (140 mcg/kg/min) MFR ≤ 2.0
Li et al.17 Rest/stress MIBI N.A. 555 925 Adenosine (140 mcg/kg/min) MFR < 2.12
Miyagawa et al.12 Stress/rest Tc MPI 6.34 3MBq/kg 9 MBq/kg Adenosine (160 mcg/kg/min) MFR < 1.3
Nkoulou et al.18 Stress/rest Tc MPI N.A. ~1000 330±33 Adenosine (140 mcg/kg/min) MFR <1.26
Shiraishi et al.19 Stress/rest Tl MPI 15.5 50–60 50–60 Adenosine 0.120 mg/kg/min) MFR ≤ 1.5
Shiraishi et al.20 Stress/rest Tl MPI 15.5 50–60 50–60 Adenosine 0.120 mg/kg/min) MFR < 2.66
Zavadovsky et al.11 Rest/stress Tc MPI 8.02 3MBq/kg 9MBq/kg Adenosine (140 mcg/kg/min) MFR ≤ 1.48
Authors CZT-SPECT Comparator Definition of comparator cut-off value Time between tests Analysis Software
Agostini et al.10 D-SPECT 15O-Water PET MFR < 2 30 d

Corridor4DM (INVIA

Medical Imaging Solutions)

Bouallègue et al.15 Discovery NM 530c FFR FFR ≤ 0.8 4 w In-house software
Han et al.16 Discovery NM 530c FFR FFR ≤ 0.8 90 d

Corridor4DM (INVIA

Medical Imaging Solutions)

Li et al.17 D-SPECT FFR FFR < 0.8 3 days Corridor4DM (INVIA Medical Imaging Solutions)
Miyagawa et al.12 Discovery NM 530c FFR FFR < 0.8 2 m NA
Nkoulou et al.18 Discovery NM 530c 13NH3-ammonia PET MFR < 2.0 12 weeks PMOD software package (version 3.1; PMOD Technologies Ltd.)
Shiraishi et al.19 Discovery NM 530c FFR FFR < 0.8 30 d In-house software
Shiraishi et al.20 Discovery NM 530c FFR FFR < 0.8 90 d AZE Virtual Place Hayabusa (AZE, Ltd., Tokyo, Japan)
Zavadovsky et al.11 Discovery NM 570c FFR FFR ≤ 0.8 7 d

Corridor4DM (INVIA

Medical Imaging Solutions)

CZT-SPECT, cadmium-zinc-telluride single-photon emission tomography; FFR, fractional flow reserve; MFR, myocardial flow rate; MPI, myocardial perfusion imaging; NA, not available; PET, positron emission tomography; Tc, technetium-99m; Tl, Thallium-201