Table 4.
Evaluation of the precision of the automated bioelectrical impedancemeter Tanita® BC-532 to assess the total body fat, compared to the dual-energy x-ray absorptiometry: residual values
| Boys (N = 93) TBF DEXA (%) = [0.617 × TBF BIA Tanita® BC-532 (%)] + 26.24 |
Girls (N = 104) TBF DEXA (%) = [0.648 × TBF BIA Tanita® BC-532 (%)] + 22.17 |
|
|---|---|---|
| Residuals (%) | ||
| Mean ± SD | − 9.92 ± 6.30 | −9.36 ± 5.59 |
| Min | −24.20 | − 24.00 |
| Max | 13.60 | 15.40 |
| Residuals (absolute value) | ||
| Mean ± SD | 10.52 ± 5.22 | 9.96 ± 4.40 |
| Max | 24.20 | 24 |
TBF DEXA (%): total body fat percentage assessed with dual-energy x-ray absorptiometry
TBF BIA Tanita® BC-532 (%): total body fat percentage assessed with the bioelectrical impedance analyser Tanita® BC-532
Residuals (%) = [TBF BIA BIA Tanita® BC-532 (%)-TBF DEXA (%)]