Skip to main content
. 2014 Nov 15;23(1):130–135. doi: 10.1016/j.jfda.2014.06.004

Table 3.

The accuracy of the HPLC-UV method for the determination of peimine and peiminine.

Analyte Sample weight (g) Original (mg) Spiked (mg) Found (mg) Recovery (%) Mean recovery (%) RSD (%)
Peimine 0.9755 0.4841 0.5010 0.9827 99.53 94.18 3.75
0.9702 0.4815 0.5010 0.9332 90.17
0.9712 0.4820 0.5010 0.9454 92.51
0.9901 0.4914 0.5010 0.9537 92.30
0.9961 0.4944 0.5010 0.9620 93.34
Peiminine 0.9755 0.2915 0.3132 0.5741 90.25 91.56 2.11
0.9702 0.2899 0.3132 0.5820 93.25
0.9712 0.2902 0.3132 0.5728 90.22
0.9901 0.2958 0.3132 0.5808 90.99
0.9961 0.2976 0.3132 0.5936 94.49

HPLC-UV = high performance liquid chromatography-ultraviolet detection; RSD = relative standard deviation.