Skip to main content
. 2022 Aug 4;12:13386. doi: 10.1038/s41598-022-16961-2

Table 1.

Evaluation of contrast ratio (CR), contrast-to-noise ratio (CNR) and generalized contrast-to-noise ratio (gCNR) after imaging with the central array only or with all three arrays on the commercial phantom.

1 array 3 arrays 3 arrays + SVD
Cyst 55 mm
Contrast (dB) − 25.5 − 22.7 − 24.2
CNR 1.80 1.73 1.78
gCNR 0.96 0.96 0.96
Cyst 80 mm
Contrast (dB) − 25.2 − 22.9 − 24.0
CNR 1.79 1.79 1.81
gCNR 0.98 0.97 0.97
Cyst 110 mm
Contrast (dB) − 12.2 − 15.6 − 16.1
CNR 1.27 1.65 1.69
gCNR 0.69 0.88 0.88

Quantification is done on cyst-like structures at depths of 55, 80 and 110 mm for 91 PWs (− 30° to 30°).