. 2022 Aug 4;12:13407. doi: 10.1038/s41598-022-17319-4

Table 1.

Summary of results for the phase corrected focal spots (target at the geometric focus).

No correction InSightec
MR binary
InSightec
MRsimCT
InSightec
CT
Hydrophone
Target pressure 0.16 ± 0.05 MPa 0.20 ± 0.03 MPa 0.22 ± 0.03 MPa 0.23 ± 0.07 MPa 0.31 ± 0.03 MPa
Peak pressure 0.19 ± 0.04 MPa 0.25 ± 0.02 MPa 0.27 ± 0.05 MPa 0.27 ± 0.05 MPa 0.31 ± 0.03 MPa
Target pressure (normalized to hydrophone) 50 ± 13% 64 ± 7% 71 ± 2% 71 ± 15% 100 ± 0%
Peak pressure (normalized to hydrophone) 59 ± 10% 79 ± 6% 85 ± 10% 87 ± 10% 100 ± 0%
Target intensity (normalized to hydrophone) 26 ± 12% 42 ± 9% 50 ± 3% 52 ± 21% 100 ± 0%
Peak intensity (normalized to hydrophone) 36 ± 12% 63 ± 10% 73 ± 17% 76 ± 17% 100 ± 0%
Positioning error 1.05 ± 0.34 mm 0.80 ± 0.41 mm 0.63 ± 0.36 mm 0.72 ± 0.47 mm 0.00 ± 0.00 mm
Focal spot volume 12.48 ± 6.71 mm3 7.53 ± 1.70 mm3 6.48 ± 1.82 mm3 5.26 ± 1.86 mm3 4.48 ± 0.67 mm3