Skip to main content
. 2022 Jul 22;16:915302. doi: 10.3389/fnins.2022.915302

FIGURE 7.

FIGURE 7

There was a strong linear correlation between apical insertion angle and cochlear diameter for both perimodiolar and lateral wall electrodes. The dotted lines highlight the apical insertion angle and cochlear diameter where the electrode sweep results in a maximum amplitude CM at a more basal electrode than the most-apical one for 500-Hz stimulus frequency. As highlighted in gray, a cochlear diameter less than 9.5 mm and an apical insertion angle greater than 350 degrees may result in the most-apical electrode crossing the CF place for 500-Hz.