TABLE 2.

Comparison of multifrequency electrocochleography insertion trajectory parameters using 250- and 500-Hz acoustic stimulus and univariate linear regression with low-frequency pure tone average (125, 250, and 500-Hz) threshold shift from preoperative to 1-month post-activation for all insertions (both perimodiolar and lateral wall electrodes).

250-Hz
500-Hz
Mean ± STD or N (%) r p Mean ± STD or N (%) r p
 Number of Cochlear Microphonic (CM) Amplitude Drops (>2 μV) 2.2 ± 2.3 0.39 0.018 2.1 ± 1.4 0.34 0.05
 Largest CM Drop (dB re: 1 μV) 2.3 ± 3.7 0.52 0.002 9.3 ± 6.0 0.39 0.033
 Starting CM Amplitude (μV) 11.7 ± 11.7 0.161 0.359 9.1 ± 14.9 0.266 0.117
 Final CM Amplitude/Max CM During Insertion * 100% 81.0 ± 20.6 0.023 0.897 70.0 ± 26.5 0.092 0.593
Type of Insertion Pattern Type A 26 (76.5) 0.112 0.529 11 (32.4) 0.207 0.225
Type B 6 (17.6) 10 (29.4)
Type C 2 (5.9) 13 (38.2)