Table 2.

Outcome Measurement

Outcome Back Plate Marking Method (n=20) Standard Method (n=20) p-value
Median (IQR) Median (IQR)
(Min, Max) (Min, Max)
Time to deploy back platea (s) 15.30 (2.70) 14.50 (3.51) 0.40
(12.00, 22.50) (9.00, 23.00)
Time to deploy upper part of device (s) 16 (12.38, 20.12) 18.75 (17, 21.25) 0.03
(9.00, 33.50) (12.00, 38.50)
Total time for deploy upper part of deviceb (s) 16 (12.38, 20.50) 21.00 (18, 30.12) <0.01
(9.00, 34.00) (12.00, 50.50)
Total pause for mechanical device deployment (s) 31.5 (27.62, 35.88) 38.75 (31.12, 43.12) 0.03
(21.00, 53.00) (26.00, 73.50)
Interruption for mechanical CPR device application as proportion of total “hand off” time 43.08% 49.18% 0.02
(38.53%, 47.51%) (43.27%, 51.83%)
(10.21%, 56.38%) (8.06%, 63.36%)
Longest chest compression interruption (s) 18 (15.62, 20.50) 18.75 (17, 21.25) 0.11
(12.00, 33.50) (14.00, 38.50)
Duration of all compression interruptions (s) 70.5 (65.38, 82.50) 82.75 (70.50, 88.12) 0.20
(61.50, 400.50) (58.50, 416.00)
Compression fractiona 77.85% (2.13) 76.91% (3.06) 0.19
(73.41%, 81.68%) (69.83%, 82.16%)

Notes: All data are presented as median unless indicated otherwise.aMean, bIncluding the time spent adjusting the suction cup after starting mechanical CPR.

Abbreviations: IQR, interquartile range; SD, standard deviation.