Technical parameters of commercial immunoassay-based multiplex proteomics platforms

Platform Vendor Multi-analyte detection Type of biological matrix Sample volume requirement Sensitivity Dynamic range Median intra-assay CV (%)
Antibody arrays See Table 1 Up to 1300 Serum, plasma, tissue/cell lysate 10 μl–1.5 ml sub pg/ml–ng/ml 3- to 4-log 15–50%
Electrochemiluminescence (ECL) multiarray Meso Scale Discovery 71 Serum, plasma, cell culture supernatants, PBMCsa, urine 10–25 μl sub fg/ml–pg/ml 5-log <20%
Proximity extension assay (PEA) Olink Proteomics 1196 Serum, plasma, tissue/cell lysate, CSFb, urine, saliva, etcc 13 μl fg/ml–pg/ml 10-log 8%
Microsphere bead capture assay Luminex 500 Serum, plasma 13 μl 1–10 pg/ml 3- to 4-log 10%
Single-molecule array (Simoa) Quanterix 12 Serum, plasma, tissue/cell lysate 13 μl sub fg/ml 4-log 10%
Slow off-rate modified aptamer (SOMAmer) assay SomaLogic 5000 Serum, plasma, tissue/cell lysate, CSFb 50 μl 100 fg/ml 8-log 4.6%
Encoded gel microparticle capture assay Abcam 75 Serum, plasma 12.5 μl 500 fg/ml 4- to 5-log <10%
Microsphere bead chip-based assay AYOXXA 12 Serum 3 μl sub pg/ml 3- to 4-log <10%
a

Peripheral blood mononuclear cell.

b

Cerebrospinal fluid.

c

A variety of other sample types are compatible such as fine needle biopsies, micro-dialysis fluid, dried blood spots, synovial fluid, plaque extract, and saliva.