Skip to main content
. 2022 Aug 8;22:139. doi: 10.1186/s12880-022-00864-9

Table 1.

MR imaging protocol

Axial T1WI Axial T2WI Sagittal T2WI Axial oblique T2WI Axial DWI* Axial T1WI postcontrast
GE signa excite HD 3.0T
 Technique FSE FS FSE FSE FSE SS-EPI LAVA
 TR (ms)/TE (ms) 620/8.2 5900/121 4920/139.1 4900/131.5 4400/64.3 4.1/1.8
 Field of view (cm) 38.0 34.0 30.0 22.0 34.0 35.0
 Matrix (phase × frequency) 320 × 224 320 × 256 320 × 256 320 × 256 256 × 256 350 × 350
 Slice thickness (mm) 5 5 4 3 5 1
 Slice gap 1 1 0.4 0 1 0
 b value(s/mm2) 0, 800
GE Discovery HD750 3.0T
 Technique LAVA-Flex FS FSE FSE FSE SS-EPI LAVA
 TR (ms)/TE (ms) 4.2/1.3 5100/106.6 5620/135.6 5500/102.0 5500/62.7 7.9/4.1
 Field of view (cm) 34.0 34.0 30.0 22.0 34.0 35.0
 Matrix (phase × frequency) 320 × 224 320 × 256 320 × 256 320 × 256 128 × 128 350 × 350
 Slice thickness (mm) 3 5 4 3 5 1
 Slice gap 0 1 0.4 0 1 0
 b value (s/mm2) 0, 800

T1WI, T1-weighted imaging; T2WI, T2-weighted imaging; FS, fat suppression; FSE, fast-recovery fast spin echo; DWI, diffusion-weighted imaging; SS-EPI, single-shot echo-planar imaging; LAVA-Flex, liver acquisition with volume acceleration-flexible; GRE, gradient echo; TR, repetition time; TE, echo time

*ADC maps were calculated voxel by voxel with the monoexponential model using formula: ADC=lnS0/S800/b800-b0 where S800 and S0 are the signal intensities with and without a diffusion gradient, respectively