TABLE 3.

Number of plans that met clinical dose criteria based on RTOG 1016 for the original clinical plans and the autoplans evaluated in this study

Number of clinical/autoplans that met clinical dose recommendations
Structure Constraint Clinic Baseline Versa HD Halcyon 2100 v15.6
Spinal cord Dmax < 45 Gy 50 50 50 50 50
Brainstem Dmax < 54 Gy 48 50 50 50 50
Ipisilateral parotid Dmean < 26 Gy 29 24 21 25 24
V30Gy < 50% 40 42 40 41 42
Contralateral parotid Dmean < 26 Gy 44 41 41 40 40
V30Gy < 50% 48 47 46 46 46
Cochleae Dmax < 35 Gy 45 40 40 39 40
Optic chiasm Dmax < 54 Gy 49 50 50 50 50
Optic nerves Dmax < 54 Gy 49 50 50 50 50
Lenses Dmax < 7 Gy 44 45 46 47 45
Eyes Dmax < 35 Gy 45 45 45 45 45
Brain Dmax < 54 Gy 34 36 37 37 36
High‐dose PTV V1cc < 110% 48 49 50 48 50
V1cc < 117% 50 50 50 50 50
V95% > 95% 50 50 50 50 50
Intermediate‐dose PTV V95% > 100% 34 16 9 10 13
V95% > 80% 50 50 50 50 50
Low‐dose PTV* V95% > 100% 30 30 15 15 21
  V95% > 80% 47 47 47 47 47

Abbreviations: D max, maximum dose; D mean, mean dose; PTV, planning target volume; RTOG, Radiation Therapy Oncology Group.

*

Three patients were treated with only two PTVs.