TABLE 4.

Low‐contrast detectability measurements and evaluation with American College of Radiology (ACR) and TG‐284 quality assurance (QA) tolerances

Image series Slice number Total number of spokes ACR a TG‐284 b
ACR T1 8‐11 40 PASS PASS
ACR T2 8‐11 38 PASS FAIL
a

ACR passing criteria for LCD were 37 or more total spokes. 2

b

TG‐284 LCD passing criteria for a 3.0 T scanner were 40 total spokes. 2