. 2022 Jul 18;23(8):e13730. doi: 10.1002/acm2.13730

TABLE 5.

Transmitter gain (TG) measurement and evaluation with American College of Radiology (ACR) and TG‐284 quality assurance (QA) tolerances

Baseline TG Measurement (dB) Annual TG Measurement (dB) Relative Difference (%) Absolute Difference (dB) ACR a TG‐284 b
0.8252 0.8668 5.0 % 0.0416 PASS PASS
a

ACR transmitter gain tolerance was ±5.0 dB from the baseline. 2

b

TG‐284 transmitter gain tolerance was ±5 % from the baseline. 3