TABLE 1.
Hardware | |
---|---|
Source‐isocenter distance | 85.0 cm |
Bore diameter | 85.0 cm |
Gantry rotation speed | 60 rpm |
Linear accelerator | |
---|---|
Photon beam energy | 6 MV flattening‐filter‐free |
Dose rate | 850 MU/min |
Jaws | Two sets situated above and below the MLCs. Capable of 1 or 2 cm field widths in IEC‐Y |
MLCs | 64 binary, double‐focused, width of 6.25 mm at isocenter, maximum field width of 40 cm in IEC‐X |
Firing positions | 50 (i.e., every 7.2 degrees) |
Beam delivery modes |
IMRT (1 pass)/SBRT (4 passes)/BgRT (4 passes) Axial delivery in 2.1 mm table increments |
TPS | |
---|---|
Optimization algorithm | Accelerated proximal gradient based on FISTA algorithm |
Dose calculation algorithm | Collapsed cone convolution (CCC) superposition |
Dose calculation grid size | 2.1 mm |
Tongue‐and‐groove | Included in beam model |
kVCT | |
---|---|
Tube current | 45‐400 mA |
Tube voltage | 120 kV |
Imaging geometry | Fan beam |
Acquisition mode | Helical with pitch = 0.22, 0.5, or 1.33 |
FOV | 50 cm |
Maximum scan length | 95 cm |
Slice thickness | 1.25 mm |
kVCT isocenter distance from treatment isocenter | 38.6 cm |
PET BgRT | |
---|---|
FOV (transverse) | 50 cm |
FOV (longitudinal) | 5.0 cm |
PET spatial resolution | 4.0 × 4.0 × 2.1 mm3 |