TABLE 2.

A summary of the non‐dosimetric tests performed on the RefleXion X1 TPS

Image Input Tests
1.1 DICOM CT and RTStruct data import
1.2 Integrity of text information transfer
1.3 Image geometry (number of pixels, slices, slice thickness)
1.4 Geometric location and orientation of scan
1.5 CT density curve
1.6 Transfer of CT, RTStruct, and RTDose between RefleXion, Eclipse, and MIM
Anatomical structure tests
2.1 Structure attributes (name, type, color)
2.2 Relative electron density definition (overrides)
2.3 Couch structure and Body structure
2.4 Structures created from contours (Boolean operators)
2.5 Structures created from non‐axial contours
2.6 Integrity of imported structures from Eclipse and MIM
Image use and display tests
3.1 Window/Level setting
3.2 Geometric accuracy of CT slices
3.3 ROI analysis (HU value within a ROI)
3.4 Tools in RefleXion tool panel
3.5 Position measurements
Plan setup tests
4.1 Target localization position
4.2 Imaging protocols and extent
4.3 Safe zone (collision)
Dose display tests
5.1 Static dose points
5.2 Isodose surfaces and dose cloud
5.3 DVH tests: dose min, max, mean, and DVH plotting
Plan and treatment delivery report tests
6.1 Plan report: patient info, plan overview, treatment info
6.2 Plan report: DVH parameters, isodose display
6.3 Treatment delivery printout

The test number is shown in the first column and a short description is given in the second column.