TABLE 2.
Image Input Tests | |
---|---|
1.1 | DICOM CT and RTStruct data import |
1.2 | Integrity of text information transfer |
1.3 | Image geometry (number of pixels, slices, slice thickness) |
1.4 | Geometric location and orientation of scan |
1.5 | CT density curve |
1.6 | Transfer of CT, RTStruct, and RTDose between RefleXion, Eclipse, and MIM |
Anatomical structure tests | |
---|---|
2.1 | Structure attributes (name, type, color) |
2.2 | Relative electron density definition (overrides) |
2.3 | Couch structure and Body structure |
2.4 | Structures created from contours (Boolean operators) |
2.5 | Structures created from non‐axial contours |
2.6 | Integrity of imported structures from Eclipse and MIM |
Image use and display tests | |
---|---|
3.1 | Window/Level setting |
3.2 | Geometric accuracy of CT slices |
3.3 | ROI analysis (HU value within a ROI) |
3.4 | Tools in RefleXion tool panel |
3.5 | Position measurements |
Plan setup tests | |
---|---|
4.1 | Target localization position |
4.2 | Imaging protocols and extent |
4.3 | Safe zone (collision) |
Dose display tests | |
---|---|
5.1 | Static dose points |
5.2 | Isodose surfaces and dose cloud |
5.3 | DVH tests: dose min, max, mean, and DVH plotting |
Plan and treatment delivery report tests | |
---|---|
6.1 | Plan report: patient info, plan overview, treatment info |
6.2 | Plan report: DVH parameters, isodose display |
6.3 | Treatment delivery printout |
The test number is shown in the first column and a short description is given in the second column.