Settings of dose parameters used for proton optimization and planning.

Spot filtering settings
Iterations before spit filtering 40
Min spot meterset (MU/fx) 0.135
Max spot meterset (MU/fx) 42
Meterset limit margin (%) 5
Proton plan optimization
ROI Description Weight
PTV_TMI Min dose 1,100 cGy (RBE) 250
PTV_TMI Max dose 1,440 cGy (RBE) 1,000
Airway Max dose 650 cGy (RBE) 100
Bladder Max dose 650 cGy (RBE) 100
Bowel Max dose 650 cGy (RBE) 100
Brain Max dose 650 cGy (RBE) 100
Esophagus Max dose 650 cGy (RBE) 100
Heart Max dose 400 cGy (RBE) 100
LT optic nerve Max dose 650 cGy (RBE) 100
LT orbit Max dose 650 cGy (RBE) 100
Parotids Max dose 650 cGy (RBE) 100
RT optic nerve Max dose 650 cGy (RBE) 100
RT orbit Max dose 650 cGy (RBE) 100
Spleen Min dose 1,200 cGy (RBE) 100
Stomach Max dose 650 cGy (RBE) 100
Thyroid Max dose 650 cGy (RBE) 100
Total lung Max dose 650 cGy (RBE) 250
Total kidneys Max dose 750 cGy (RBE) 100

ROI, region of interest; PTV, planning target volume; TMI, total marrow irradiation; RBE, relative biological effectiveness.