Skip to main content
. 2022 Jun 1;35(2):ivac149. doi: 10.1093/icvts/ivac149

Figure 3:

Figure 3:

Intraoperative picture of the incised bend relief during reoperation for a patient with a compressed outflow graft obstruction. Note the jelly-like substance under the bend relief on top of the Dacron graft.