Skip to main content
. Author manuscript; available in PMC: 2023 Jul 12.
Published in final edited form as: IEEE Trans Biomed Circuits Syst. 2022 Jul 12;16(3):395–408. doi: 10.1109/TBCAS.2022.3175926

Fig. 11.

Fig. 11.

Power consumption comparison for the two hardware accelerators. (Left) SBP feature extraction with CAR from 93 channels at 2kSps. M0 controls the three SPI blocks and calculates referenced SBP with a 31.5 MHz clock at 1.35 V, while the SBPU requires a 2.9 MHz clock at 0.63 V. (Right) Continuous SSKF inference from 93 channels of SBP data every 50 ms. M0 calculates SSKF updates with a 240 kHz clock at 0.63 V or a 484 kHz clock at 0.63 V for 1D or 2D, respectively. MAU requires a 67.8 kHz clock at 0.63 V.