Table 2.
Platinum– DEXIS | GXS-700– Gendex | RVG 6100– Carestream (Kodak) | ProSensor– Planmeca | CDR Elite– Schick | XIOS Plus– Sirona | |
---|---|---|---|---|---|---|
GXS-700– Gendex | 0.67 | |||||
0.094 | ||||||
0.475−0.83 | ||||||
RVG 6100– Carestream (Kodak) | 0.8279 | 0.595 | ||||
0.0756 | 0.104 | |||||
0.642−0.937* | 0.389−0.778 | |||||
ProSensor– Planmeca | 0.6315 | 0.463 | 0.36 | |||
0.097 | 0.101 | 0.100 | ||||
0.433−0.80 | 0.277−0.658 | 0.189−0.566 | ||||
CDR Elite– Schick | 0.6041 | 0.373 | 0.267 | 0.41 | ||
0.097 | 0.098 | 0.086 | 0.102 | |||
0.408−0.776 | 0.202−0.574 | 0.128−0.456* | 0.232−0.615 | |||
XIOS Plus– Sirona | 0.699 | 0.46 | 0.38 | 0.509 | 0.598 | |
0.0895 | 0.100 | 0.097 | 0.101 | 0.098 | ||
0.507−0.847* | 0.278−0.657 | 0.214−0.581 | 0.318−0.699 | 0.403−0.772 | ||
XDR– Cyber Medical Imaging | 0.478 | 0.347 | 0.246 | 0.383 | 0.416 | 0.343 |
0.103 | 0.096 | 0.094 | 0.098 | 0.101 | 0.096 | |
0.287−0.674 | 0.183−0.546 | 0.102−0.460* | 0.250−1.446 | 0.236−0.616 | 0.180−0.543 |
Each entry lists the area under the curve (AUC), followed by the standard deviation (SD) and then by the 95% confidence interval. *Indicates significant differences between pairs (P < 0.05). XDR (Cyber Medical Imaging), RVG 6100 (Carestream Dental LLC), Platinum (DEXIS LLC), CDR Elite (Schick Technologies), ProSensor (Planmeca), EVA (ImageWorks), XIOS Plus (Sirona) and Gendex GXS-700 (Gendex Dental Systems).