Skip to main content
. 2022 Jul 8;6:246. Originally published 2021 Sep 24. [Version 3] doi: 10.12688/wellcomeopenres.17202.3

Table 1. Linearity, accuracy (%) and precision (%) for the quantification of efavirenz in cervicovaginal swab (CVS).

SD=standard deviation.

Nominal
concentration
(ng/mL)
Measured concentration (inter-
day)
Measured concentration (intra-
day)
mean SD precision
(%CV)
accuracy
(%)
mean SD precision
(%CV)
accuracy
(%)
25 25.7 1.31 5.10 102.7 25.6 1.63 6.36 102.3
50 47.2 4.44 9.40 94.5 48.0 5.32 11.1 96.0
100 91.9 6.61 7.19 91.9 91.8 4.75 5.18 91.8
200 193 11.6 6.00 96.4 195 17.3 8.89 97.3
500 498 41.9 8.41 99.5 480 57.0 11.9 96.0
1000 1010 70.9 7.00 101.4 1030 93.1 8.99 103.5
2000 2030 120 5.89 101.4 2005 112.3 5.60 100.3
4000 3990 208 5.22 99.8 4050 138.3 3.41 101.3
7500 7780 402 5.17 103.7 7940 457 5.75 105.9
10000 10400 768 7.39 104.0 10530 1020 9.68 105.3
25 (LLOQ) 24.6 2.01 8.14 98.5 24.2 1.86 7.69 96.9
75 (LQC) 75.0 6.33 8.44 99.9 74.3 4.38 5.90 99.1
4500 (MQC) 4640 457 9.85 102.7 4740 496 10.5 105.3
8500 (HQC) 8440 983 11.6 105.2 8470 1260 14.9 99.6