Table 1.

FLIP protocol by catheter type

Catheter configuration
 Balloon length EF-325: 8 cm EF-322: 16 cm
 Sensors 16 sensors spaced 0.5 cm apart 16 sensors spaced 1 cm apart
Protocol
 Baseline placement Balloon should span the EGJ with the EGJ waist in the mid-balloon length Balloon should span the EGJ with 2–3 sensors distal to the EGJ
 Baseline fill volume 20 mL 30 mL
 Baseline wait time 15 s 15 s
 Balloon fill protocol Fill to 30 mL, 40 mL, 50 mL Fill to 40 mL, 50 mL, 60 mL, 70 mLa
 Wait time at each fill 30 s 60 s
Measurements
EGJ-DIb Diameterb Intrabag pressureb EGJ-DIb Diameterb Intrabag pressureb Presence of RACs Presence of RRCs Presence of any contractility

EGJ, esophagogastric junction; EGJ-DI, EGJ-distensibility index; FLIP, functional lumen imaging probe; RAC, repetitive anterograde contraction; RRC, repetitive retrograde contraction.

a

Fill to 70 mL if intrabag pressures do not reach greater than 15–20 mm Hg or outflow obstruction is suspected.

b

Measured when the narrowest luminal diameter achieves maximal diameter.