Table 1.
FLIP protocol by catheter type
| Catheter configuration | ||
| Balloon length | EF-325: 8 cm | EF-322: 16 cm |
| Sensors | 16 sensors spaced 0.5 cm apart | 16 sensors spaced 1 cm apart |
| Protocol | ||
| Baseline placement | Balloon should span the EGJ with the EGJ waist in the mid-balloon length | Balloon should span the EGJ with 2–3 sensors distal to the EGJ |
| Baseline fill volume | 20 mL | 30 mL |
| Baseline wait time | 15 s | 15 s |
| Balloon fill protocol | Fill to 30 mL, 40 mL, 50 mL | Fill to 40 mL, 50 mL, 60 mL, 70 mLa |
| Wait time at each fill | 30 s | 60 s |
| Measurements | ||
| EGJ-DIb Diameterb Intrabag pressureb | EGJ-DIb Diameterb Intrabag pressureb Presence of RACs Presence of RRCs Presence of any contractility | |
EGJ, esophagogastric junction; EGJ-DI, EGJ-distensibility index; FLIP, functional lumen imaging probe; RAC, repetitive anterograde contraction; RRC, repetitive retrograde contraction.
Fill to 70 mL if intrabag pressures do not reach greater than 15–20 mm Hg or outflow obstruction is suspected.
Measured when the narrowest luminal diameter achieves maximal diameter.