Skip to main content
. 2022 Apr 7;32(9):6215–6229. doi: 10.1007/s00330-022-08753-z

Table 2.

MRI acquisition parameters

Sequence FS Orientation TA Acquired voxel size Slices FOV TE TR FA AV C PAT TF
3T PD TSES FS Sagittal 3:11 0.67 × 0.47 × 3.0 30 150 × 150 44 3790 150 2 1 3 7
PD TSEDL FS Sagittal 1:33 0.69 × 0.47 × 3.0 30 150 × 150 41 3580 150 1 1 3 7
PD TSES FS Coronal 3:11 0.67 × 0.47 × 3.0 30 150 × 150 44 3790 150 2 1 3 7
PD TSEDL FS Coronal 1:33 0.69 × 0.47 × 3.0 30 150 × 150 41 3580 150 1 1 3 7
PD TSES FS Axial 3:11 0.67 × 0.47 × 3.0 30 150 × 150 44 3790 150 2 1 3 7
PD TSEDL FS Axial 1:33 0.69 × 0.47 × 3.0 30 150 × 150 41 3580 150 1 1 3 7
T1w TSES Coronal 2:23 0.42 × 0.33 × 3.0 30 150 × 150 10 566 150 1 1 2 3
T1w TSEDL Coronal 1:32 0.42 × 0.33 × 3.0 30 150 × 150 12 448 150 1 3 4 3
1.5T PD TSES FS Sagittal 3:12 0.67 × 0.47 × 3.0 30 150 × 150 42 3800 150 2 1 3 7
PD TSEDL FS Sagittal 1:18 0.65 × 0.47 × 3.0 30 150 × 150 40 3630 150 1 1 4 7
PD TSES FS Coronal 3:12 0.67 × 0.47 × 3.0 30 150 × 150 42 3800 150 2 1 3 7
PD TSEDL FS Coronal 1:18 0.65 × 0.47 × 3.0 30 150 × 150 40 3630 150 1 1 4 7
PD TSES FS Axial 1:56 0.67 × 0.47 × 3.0 30 150 × 150 16 3800 150 2 1 2 7
PD TSEDL FS Axial 1:00 0.74 × 0.47 × 3.0 30 150 × 150 15 3420 150 1 1 4 7
T1w TSES Coronal 2:13 0.42 × 0.33 × 3.0 30 150 × 150 9 527 150 1 2 2 3
T1w TSEDL Coronal 1:14 0.47 × 0.33 × 3.0 30 150 × 150 10 593 150 1 2 4 3

Acquired voxel size in mm3

TA time of acquisition, FOV field of view (mm2), TE/TR echo time/repetition time (ms); FA flip angle (degree), AV averages, C concatenations, PAT parallel acquisition technique, TF turbo factor, TSE turbo spin echo, PD proton density, T1w T1-weighted, FS fat saturation