Ablation Settings

Setting Number Duration
(s)
Power
(W)
Flowrate
(mL/min)
Saline Temperature
(°C)
Target Lesion
Volume* (cc)
1 30 20 10 60 0.50
2 40 25 10 60 1.10
3 55 35 10 60 2.30
4 65 45 10 60 3.60
5 80 50 10 60 5.00

S = Seconds; W = power in watts; ml/min = flow rate in milliliters/minutes; °C = degrees in Centigrade; cc = cubic centimeters anticipated with settings 1–5.

*

from references [27, 36]