Table 2.

Sterilization conditions for frozen mice based on EN ISO 17665-1.

Conditions Temperature Duration Pressure
Advanced sterilization 1.0 Kelvin
Fractionated prevacuum stages Stage 1 Stage 2 Stage 3 Stage 4
Number of repetitions 1 1 4 4
Pressure steam injection 600 mbar /8.7 psi 1000 mbar /14.5 psi 1900 mbar /27.6 psi 2500 mbar /36.3 psi
Pressure steam flush II 90 mbar /1.3 psi 90 mbar /1.3 psi 1300 mbar /18.9 psi 2000 mbar /29 psi
Flow steam flush II 100% 100% 100% 100%
Pressure evacuation 75 mbar /1.1 psi 200 mbar /2.9 psi 1300 mbar /18.9 psi 2000 mbar /29 psi
Pressure steam flush I 90 mbar /1.3 psi 215 mbar /3.1 psi 1300 mbar /18.9 psi 2000 mbar /29 psi
Time steam flush I 60 s 60 s
Flow steam flush I 10% 10% 100% 100%
Sterilization 127 °C 20 min 2500 mbar / 36.3 psi
Drying /Airflow . 3 min 800 mbar / 11.6 psi
Ventilation pressure 1100 mbar / 16 psi

I: Evacuation and flow pressure used for lower pressure levels

II: Evacuation and flow pressure used for higher pressure levels

Advanced sterilization: This value is added to the set the sterilization temperature

Flow steam flush I: Opening percentage of the vacuum valve

Flow steam flush II: Opening percentage of the vacuum valve

Number of repetitions: This value indicates how often the process element (stage) is repeated with an identical procedure

Pressure steam injection: Set pressure for steam injection

Pressure evacuation: Set pressure for evacuation

Pressure steam flush II: Set pressure for steam flush II

Pressure steam flush I: Set pressure for steam flush I

Time steam flush I: Holding time at steam flush pressure I