TABLE 2.

Calculated maximum charge for each electrode (Troyk et al., 2007; Kane et al., 2013).

Parameter Value
Anticipated duration 5 years
Daily electrode use 4 h
Expected clinical testing 3 days/week 3 weeks/month
Clinical frequency 100 pulses/second
Duty cycle 50%
Total number of pulses 3.888 × 108 pulses
Maximum current 80 μA
Clinical pulse width 200 μs
Maximum charge/Pulse 16 nC
Calculated total charge 6.22 Coulombs