. 2022 Aug 18;14:1427–1436. doi: 10.2147/NSS.S367400

Table 3.

The Performance of Reported PPG Based Oximetry for Sleep Apnea Testing in Detecting Suspected OSA

Study Portable Monitoring PPG Technology N AHI Threshold Sensitivity (%) Specificity (%) PPV (%) NPV (%)
Gu et al, 202017 Belun Ring rPPG 50 AHI≥5 95 29 89.1 50
AHI≥15 85 87 88.5 65.5
Massie F et al, 201819 NightOwl rPPG 101 AHI>≥5 92 100 56 100
AHI≥15 94 87 89 92
Papini G B et al, 202020 Wrist-worn oximetry rPPG 502 AHI≥5 72 71 84 NR
AHI≥15 59 90 75 NR
Ayas N et al, 200321 Watch-PAT tPPG 30 AHI≥15 93.3 73.3 NR NR
Romem A et al, 201418 Morpheus Ox tPPG 65 AHI≥5 80 86 93 68
AHI≥15 70 91 80 85
Our data Circul rPPG 207 AHI≥5 87 83 96 57
AHI≥15 66 96 97 62

Abbreviations: PPG, photoplethysmography; rPPG, reflective photoplethysmography; tPPG, transmissive photoplethysmography; PPV, positive predictive value; NPV, negative predictive value; AHI, apnea-hypopnea index; NR, not reported.