Table 3.
The Performance of Reported PPG Based Oximetry for Sleep Apnea Testing in Detecting Suspected OSA
Study | Portable Monitoring | PPG Technology | N | AHI Threshold | Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) |
---|---|---|---|---|---|---|---|---|
Gu et al, 202017 | Belun Ring | rPPG | 50 | AHI≥5 | 95 | 29 | 89.1 | 50 |
AHI≥15 | 85 | 87 | 88.5 | 65.5 | ||||
Massie F et al, 201819 | NightOwl | rPPG | 101 | AHI>≥5 | 92 | 100 | 56 | 100 |
AHI≥15 | 94 | 87 | 89 | 92 | ||||
Papini G B et al, 202020 | Wrist-worn oximetry | rPPG | 502 | AHI≥5 | 72 | 71 | 84 | NR |
AHI≥15 | 59 | 90 | 75 | NR | ||||
Ayas N et al, 200321 | Watch-PAT | tPPG | 30 | AHI≥15 | 93.3 | 73.3 | NR | NR |
Romem A et al, 201418 | Morpheus Ox | tPPG | 65 | AHI≥5 | 80 | 86 | 93 | 68 |
AHI≥15 | 70 | 91 | 80 | 85 | ||||
Our data | Circul | rPPG | 207 | AHI≥5 | 87 | 83 | 96 | 57 |
AHI≥15 | 66 | 96 | 97 | 62 |
Abbreviations: PPG, photoplethysmography; rPPG, reflective photoplethysmography; tPPG, transmissive photoplethysmography; PPV, positive predictive value; NPV, negative predictive value; AHI, apnea-hypopnea index; NR, not reported.