Analytical results obtained for Certified Reference Material (CRM) samples along with per cent recovery.

Trace metals Reference # 7502-a (White rice powder) JSd-1 (Stream sediment)
Certified value (µg g-1) Observed value (µg g-1) Recovery (%) Certified value (µg g-1) Observed value (µg g-1) Recovery (%)
Cu 3.02 3.08 ± 0.060 102.0 22.00 22.62 ± 1.21 102.8
Zn 26.00 24.62 ± 0.041 94.7 96.50 104.22 ± 0.48 108.0
Ni 0.39 0.414 ± 0.008 106.0 7.04 7.66 ± 0.10 108.8
Fe 4.48 4.93 ± 0.071 110.0 20,010 ± 511
Mn 11.20 10.18 ± 0.085 90.9 488.9 ± 2.24
Pb 0.0043 bdl (< 0.01) 12.90 13.80 ± 4.38 107.0
Cr 0.075 0.084 ± 0.036 112.0 21.50 20.36 ± 1.22 94.7
Cd 0.548 0.522 ± 0.18 95.3 0.146 0.134 ± 0.08 91.8

bdl  below detectable limit.