Table 2.

Imaging parameters of quantitative MRI protocol.

Sequence T2-mapping 2D T2w-FSE 2D Multi-shell DTI-EPI 3D, 2-echo Dixon GRE 2D B1+ mapping
Scan time (min:sec) 4:12 4:56 0:48 0:44
Orientation Axial Axial Axial Axial
Field-of-view (FOV) (cm) 27.0 27.0 34.0 27.0
TR/TE (ms) 2400/10–80 3300/56 4.2/1.2, 2.5 11/7
Matrix size (frequency x phase) 192 × 128 90 × 90 256 × 256 64 × 64
Phase FOV factor 1.50 - - -
Slice thickness (mm), Spacing 4.0, 1.0 3.0, - 1.8, - 5.0, -
Slices 26 35 96 26
Bandwidth (kHz) ±15.6 ±250.0 ±142.9 ±15.6
Echo train length 8 51 2 -
Parallel imaging factor (phase) 2 1.5 1.5 -
GRE Flip Angle (°) - - 10 10
b-values (s/mm2) [# directions] - 0 [2]
375 [13]
750 [27]
[Total=40]
- -
Fat suppression method None Spectral-spatial pulse Dixon None

FSE: fast/turbo spin echo, DTI: diffusion tensor imaging, GRE: gradient recalled echo, TR: repetition time, TE: echo time