Mode Deviation User instruction Prosthesis setting
Stand to sit Unable to initiate sitting Cue for active hip flexion and equal weight bearing between limbs for increased load onto the prosthesis Decrease axial load threshold
Inadequate support during sitting N/A Increase or decrease knee damping to provide more or less support, respectively
Sit to stand Unable to initiate standing Cue to increase forward trunk position and load onto the prosthesis Decrease axial load threshold to initiate knee extension power
Inadequate support or power during standing Cue for equal weight bearing between limbs for increased load onto the prosthesis Increase stance phase knee stiffness for improved support and power into knee extension