Tolerance levels used in the Periodic Image Quality Test (PIQT) tests that monitor slice profile

Imaging sequence FWHM (mm) Slice integral (mm)
MS‐MESE Echo 1: 4.65–5.15 Echo 1: 4.85–5.35
Echo 2: 4.45–4.95 Echo 2: 4.65–5.15
MS‐FFE 4.75–5.25 4.90–5.40

Abbreviations: FWHM, full width at half maximum; MS‐FFE, multi‐slice fast field‐echo; MS‐MESE, multi‐slice multi‐echo spin‐echo.