. 2022 Mar 25;23(6):e13586. doi: 10.1002/acm2.13586

TABLE 5.

Mean and standard deviation of the signal‐to‐noise ratio (SNR) and uniformity over the course of the first 150 days

Imaging sequence SNR (μ ± σ) Uniformity (μ ± σ) (%)
MS‐MESE Echo 1: 69.3 ± 2.0 Echo 1: 38.6 ± 0.5
Echo 2: 51.1 ± 1.5 Echo 2: 38.6 ± 0.6
MS‐FFE 59.3 ± 1.8 38.7 ± 0.6
2D‐MESE Echo 1: 53.3 ± 2.2 Echo 1: 5.8 ± 0.3
Echo 2: 46.1 ± 2.0 Echo 2: 6.3 ± 0.3
Echo 3: 37.3 ± 1.9 Echo 3: 6.5 ± 0.4

Note: Tolerance levels for each sequence are given in Table 2.

Abbreviations: 2D‐MESE, two‐dimensional multi‐echo spin‐echo; MS‐FFE, multi‐slice fast field‐echo; MS‐MESE, multi‐slice multi‐echo spin‐echo.