TABLE 6.

Mean and standard deviation of the full width at half maximum (FWHM) and slice integral over the course of 1 year

Imaging sequence FWHM (μ ± σ) (mm) Slice integral (μ ± σ) (mm)
MS‐MESE Echo 1: 4.93 ± 0.02 Echo 1: 5.10 ± 0.04
Echo 2: 4.67 ± 0.03 Echo 2: 4.87 ± 0.05
MS‐FFE 4.97 ± 0.03 5.15 ± 0.05

Note: Tolerance levels for each sequence are given in Table 3.

Abbreviations: FWHM, full width at half maximum; MS‐FFE, multi‐slice fast field‐echo; MS‐MESE, multi‐slice multi‐echo spin‐echo.