TABLE 9.

Results of image quality assessment using American College of Radiology (ACR) testing recommendations

T1‐weighted T2‐weighted Tolerance
Geometric accuracy <1 mm <1 mm <±2 mm
Slice position accuracy 1.8 mm 1.9 mm −5 mm < x < 5 mm
Slice thickness 4.8 mm 4.7 mm 5 ± 0.7 mm
PIU 91.4% 91.8% >87.5%
PSG 0.0012 0.0014 ≤0.025
HCSR 0.9 0.9 ≤1
LCD 28 22 >9