TABLE A1.

Nominal T1 and T2 values in the National Institute of Standards and Technology (NIST)/International Society of Magnetic Resonance in Medicine (ISMRM) phantom used in this study

T1 (ms) T2 (ms)
1879.0 1044.0
1432.0 623.9
1027.0 428.3
751.3 258.4
527.0 186.1
384.1 137.0
272.3 89.52
194.5 62.82
137.8 43.84
94.7 27.28
67.0 19.24
48.14 15.44
34.35 10.05
24.16 7.79