Table 1.

Clinical validation of the COVID-19-RT-LAMP-XO by visual observation and RT-LAMP-DETR against RT-qPCR on ORF1ab gene.

Result Ct value (ORF1ab gene) RT-LAMP-XO by visual observation/RT-LAMP-DETR
Positive Negative Sum (N)
RT-qPCR Positive 0–25.0 34/34 0/0 34
25.1–30.0 19/19 0/0 19
30.1–35.0 19/19 0/0 19
35.1–40.0 16/16 0/0 16
Negative >40/Undetectable 0/0 125/125 125
Total 88/88 125/125 213
% Concordance with respect to RT-qPCR result 100 100

N denotes sample size.