. 2022 Aug 7;10(8):1485. doi: 10.3390/healthcare10081485

Table 2.

Pressure Pain Threshold (PPT) Values of Neck and Shoulder in Volunteers/Patients with and without Neck/shoulder Pain.

Authors Object of Study/Subject Areas of PPT Examination Device (Units) PPT Values in Patients/People with Neck and Shoulder Pain; Mean (±SD) PPT Values in Healthy Volunteers/Control; Mean (±SD)
Arjona Retamal JJ et al. 2021 Chronic neck pain Upper trapezius Digital algometer (kg/cm2)
(FPX 25, Wagner Ins, Greenwich, CT)
Upper trapezius: 1.35–1.56 none
Leon Hernández JV et al. 2021 Chronic neck pain Upper trapezius Digital algometer (kg/cm2)
(Wagner Ins, greenwich, CT, USA)
Uppe tripezius: 4.11–4.14 none
Stieven FF et al. 2021 Chronic neck pain Upper trapezius Digital algometer (kgf)
(FPX 25, Wagner Ins, Greenwich, CT)
Upper trapezius: 1.35–1.46 none
Oliveira AK et al. 2021 Chronic neck pain Upper trapezius Algometer (kg/cm2)
(PTR-300 model, Instrutherm, São Paulo, SP, Brazil)
Upper trapezius: 1.44–1.76 none
Grimby-Ekman A et al. 2020 Chronic neck-shoulder pain Upper trapezius Distal algometer (kpa)
(Somedic AB, Farsta, Sweden)
Upper trapezius: 376–411 Upper trapezius: 335–436
Heredia-Rizo AM et al. 2020 Neck-shoulder pain Upper trapezius Electronic pressure algometer (kpa)
(Somedic AB, Horby, Sweden)
Upper trapezius: 226.1 ± 103.2
(177.8–274.4)
Upper trapezius: 282.2 ± 109.4
Arias-Buria JL et al. 2020 Mechanical neck pain Triger point area of shoulder Mechanical algometer (kpa)
(Pain Diagnosis and Treatment Inc., New York, NY, USA)
Triger point area of shoulder: 145.4–148.9 none
Shin HJ et al. 2020 Chronic neck pain Upper trapezius
Splenius capitis
Levator scapulae
Digital algometer (kg)
(Somedic AB, Farsta, Sweden)
Upper trapezius: 2.41–2.56
Splenius capitis: 2.56–2.90
Levator scapulae: 2.07–2.38
none
Rodríguez-Huguet M et al. 2020 Mechanical neck pain Upper trapezius
Suboccipital
Algometer (N/cm2)
(Wagner Ins, Greenwich, CT, USA)
Upper trapezius: 1.46
Suboccipital: 1.20–1.22
none
Alfawaz SS et al. 2020 Non-specific
mechanical neck pain
Upper trapezius Handheld algometer (N/cm2)
(Force five™, Wagner Ins, Greenwich, CT, USA)
Upper trapezius: 4.0–4.9 none
Chatchawan U et al. 2019 Chronic tension-type headache (CTTH)
Episodic tension-type headaches (ETTH)
Head, neck, shoulder and upper back. Manual algometer (kg/cm2)
(Force Dial FDK/FDN Series Mechanical Force Gage; Wagner Ins, Greenwich, CT, USA).
Head, neck, shoulder and upper back: 0.7–1.2 none
Wang-Price S et al. 2018 Neck-shoulder pain Middle deltoid
Levator scapulae
Upper trapezius
Handheld computerized pressure algometer (kpa)
(Medoc Ltd., Ramat Yishai, Israel)
Middle deltoid: 194.7–228.6
Levator scapulae: 244.1–246.5
Upper trapezius: 167.6–204.1
Middle deltoid: 248.9–309.2
Levator scapulae: 313.7–322.0
Upper trapezius: 229.3–234.6
Murray M et al. 2017 Neck-shoulder pain among military helicopter pilots and crew Trapezius
Upper neck extensors
Handheld electronic pressure algometer (kpa)
(Type II Algometer, Somedic Production AB, Sweden)
Trapezius: 405–434
Upper neck extensors: 334–347
none
De Meulemeester KE et al. 2017 Myofascial neck/shoulder pain Upper and middle trapezius, levator scapulae, infraspinatus and supraspinatus Wagner FPX Digital Algometer (kg/cm2) Upper and middle trapezius, levator scapulae, infraspinatus and supraspinatus: 16.20–20.63 none
Wassinger CA et al. 2016 Shoulder pain. Shoulder Electronic pressure algometer (kg/cm2)
(Wagner Ins, Greenwich, CT, USA)
Shoulder: 5.67–5.73 none
Toprak Celenay S et al. 2016 Neck pain C7 and acromion at the middle point of the upper trapezius muscle Digital algometer (kg/cm2)
(JTech Medical Industries, ZEVEX Company)
Trapezius: 7.05–7.74 none
Pajediene E et al. 2015 Whiplash associated pain Upper part of the body Hand-held pressure algometer (kg/cm2)
(Pain TestTM Algometer, Wagner Force dial FDK 20)
Upper part of the body: 2.87 ± 1.21 Upper part of the body: 3.71 ± 1.69
Lopez-Lopez A et al. 2015 Chronic neck pain C2 spinous process Digital algometer (N/cm2)
(FDX 25, Wagner Ins, Greenwich, CT, USA)
C2 spinous: 1.49–1.70 none
Ge HY et al. 2014 Computer users with or without pain in the neck-shoulder and forearmregions Most painful or dominant side of the neck-shoulder region Digital algometer (kpa)
(Somedic AB, Hörby, Sweden)
Most painful or dominant side of the neck-shoulder region: 203.4–308.6 Most painful or dominant side of the neck-shoulder region: 238.7–337.1
Llamas-Ramos R et al. 2014 Chronic neck pain C7 spinous process Mechanical algometer (kpa)
(Pain Dignosis and Treatment Inc., New York, NY, USA)
C7 spinous process: 188.1 ± 49.4 none
Andersen CH et al. 2014 Chronic neck/shoulder pain Upper trapezius
Lower trapezius
Pressure algometer (kpa)
(Algometer Type 2, Somedic, Hörby, Sweden)
Upper trapezius: 277–303
Lower trapezius: 308–383
none
Casanova-Méndez A et al. 2014 Chronic neck pain Upper tarapezius
C4 spinous process
T4 spinous process
Analog pressure algometer (kg/cm2)
(Baseline®, FEI Inc., White Plains, NY, USA)
C4 spinous process: 1.96–2.01
T4 spinous process: 3.35–3.70
Upper tripezius: 2.79–3.46
none
Cagnie B et al. 2013 Office workers with mild neck and shoulder complaints Triger point in Levator scapula/Upper Trapezius/Splenius cervicis Electronic algometer (N)
(compuFET; Hoggan Health Industries, Inc., West Jordan, UT, USA)
Triger point in Levator scapula/Upper Trapezius/Splenius cervicis: 16.2–24.5 none
Yoo IG et al. 2013 Neck-shoulder pain Upper trapezius
Middle trapezius
Dolorimeter pressure algometer (lb)
(Fabrication Enterprises, White Plains, NY, USA)
Upper trapezius: 7.2 ± 1.8
Middle trapezius: 5.8 ± 1.4
Upper trapezius: 6.3 ± 2.0
Middle trapezius: 5.0 ± 1.2
Lauche R et al. 2013 Chronic neck pain Levator scapula
Trapezius upper
Semispinalis capitis
Digital algometer (kpa)
(Somedic AB, Hörby, Sweden)
Levator scapula: 273.4–343.6
Trapezius upper: 229.0–273.7
Semispinalis capitis: 178.9–219.5
none
Casanova-Méndez A et al. 2013 Chronic neck pain C4 and T4 spinous process
Trapezius
Analogue pressure algometer (kg/cm2)
(Baseline, FEI Inc., White Plains, NY, USA)
C4 spinous process: 1.96–2.01
T4 spinous process: 3.35–3.70
Trapezius: 2.79–3.46
none
Fernández-Pérez AM et al. 2012 Whiplash associated pain Articular pillar of the C5–6 zygapophyseal joints Electronic algometer (kpa)
(Somedic AB, Sweden)
Articular pillar of the C5–6 zygapophyseal joints: 139.8–158.0 Articular pillar of the C5–6 zygapophyseal joints: 205.4–212.7
Andersen LL et al. 2012 Neck-shoulder pain Upper Trapezius Electronic pressure algometer (kpa)
(Wagner Ins, greenwich, CT, USA)
Upper trapezius: 219–260 none
Andersen LL et al. 2012 Neck/shoulder pain Upper trapezius Electronic pressure algometer (kpa)
(Wagner Ins, Greenwich, CT, USA)
Trapezius: 219–260 none
Ge HY et al. 2009 Fibromyalgia syndrome Upper trapezius Electronic algometer (kpa)
(Somedic AB, Sweden)
Trapezius: 151–156 Trapezius: 151–156
Gerdle B et al. 2008 Whiplash associateddisorders Trapezius Electronic pressure algometer (kpa)
(Somedic Algometer type 2, Sollentuna, Sweden)
Trapezius, low cervical and spraspinatus: 95–130 Trapezius, low cervical and spraspinatus: 230–300
Lemming D et al. 2007 Whiplash associated pain Infraspinatus Electronic algometer (kpa)
(Somedic AB, Sweden)
Infraspinatus: 224.6–270.5 none
Ylinen J et al. 2007 Chronic neck pain Splenius capitis
Trapezius
Levator scapulae
Hand-held digital pressure algometer (N/cm2)
(Force fiveTM, Wagner Instruments, Box 1217, Greenwich, CT 06836, USA)
Splenius capitis: 38.9–39.6
Trapezius: 38.3–40.3
Levator scapulae: 60.2–60.7
none
Ylinen J et al. 2007 Chronic neck pain Strenum
Trapezius middle
Lavator scapulae
Trapezius upper
Handheld algometer (N/cm2)
(Force five™, Wagner Ins, Greenwich, CT, USA)
Strenum: 31–34
Trapezius middle: 28–31
Lavator scapulae: 43–51
Trapezius upper: 27–31
none
Ojala T et al. 2006 Neck-Shoulder Myofascial Pain Syndrome Trigger point in neck and shoulder Dolorimeter (kg/cm2)
(Pain Diagnostic, Fisher).
Trigger point in neck and shoulder: 5.1–5.3 none
Ylinen J et al. 2005 Chronic neck pain Sternum
Trapezius middle
Levator scapulae
Trapezius upper
Handheld electronic pressure algometer (N/cm2)
(Force five™, Wagner Ins, Greenwich, CT, USA)
Strenum: 31–36
Trapezius middle: 28–38
Lavator scapulae: 45–59
Trapezius upper: 28–38
none
Nabeta T et al. 2002 Neck/shoulder pain Neck
Shoulder
Pressure algometer (kg/cm2)
(Yufu-Seiki, F P Meter, with probe of 10 mm)
Neck: 1.6–1.7
Shoulder: 2.0–2.4
none
Waling K, Sundelin G et al. 2000 Work-related trapezius myalgia Trapezius upper/middle/lower Somedicw pressure algometer (kpa)
(Somedic Production AB, Sollentuna, Sweden).
Trapezius upper/middle/lower: 194–253 none
Taimela S et al. 2000 Chronic neck pain Trapezius
Levator Scapulae
Mechanical force gauge (N/cm2) Trapezius: 25.9–29.6
Levator Scapulae: 40.9–41.7
none
Otto A et al. 2019 Healthy volunteer Deltoid
Upper trapezius
Pressure algometer (kpa)
(Somedic AB, Farsta, Sweden, probe size of 1 cm2 surface area)
none Deltoid: 304.42 ± 103.42
Upper trapezius: 77.3 ± 126.90
Sacramento LS et al. 2017 Healthy Children and Young Adults Upper Trapezius
Supraspinatus
Infraspinatus
Levator Scapulae
Deltoid
Digital pressure algometer (kg/cm2)
(OE-220, ITO Physiotherapy and Rehabilitation, Ito, Japan)
none Children: Upper Trapezius: 1.4 ± 0.6
Supraspinatus: 2.0 ± 0.7
Infraspinatus: 2.2 ± 0.7
Deltoid: 2.1 ± 0.8 C5–6
joint: 1.3 ± 0.5
Adults: Upper Trapezius: 2.5 ± 0.9
Supraspinatus: 3.5 ± 1.3
Infraspinatus: 3.8 ± 1.3
Deltoid: 2.7 ± 1.1 C5–6
joint: 2.1 ± 0.7
Wytrążek M et al. 2014 Healthy volunteerwithout triger points Trapezius (upper part) n = 13
Sternocleido-mastoid n = 12
Deltoid (middle part) n = 48
Infraspinatus n = 13
Algometer (kg/cm2)
(Force Dial FDK/FDN Series Push Pull Force Gage;
Wagner Instruments, Riverside, CT, USA)
none Trapezius (upper part): 8.38–9.5
Sternocleido-mastoid: 5.92–6.33
Deltoid (middle part): 7.85–8.56
Infraspinatus: 8.32–10
Shin SJ et al. 2012 Healthy worker Upper trapezius Dolorimeter (Lb)
(Fabrication Enterprises, White Plains, NY, USA)
none Upper trapezius: 7.3–8.8
Binderup AT et al. 2010 Healthy volunteer Upper trapezius
Middle trapezius
Lower trapezius
Spinal processes
Hand-held algometer (kpa)
(Somedic® Algometer type 2, Sweden)
none Upper trapezius: 295.2 ± 95.9
Middle trapezius: 347.5 ± 103.5
Lower trapezius: 373.0 ± 121.1
Spinal processes: 369.6 ± 116.5
Saíz-Llamosas JR et al. 2009 Healthy volunteer C5-C6 zygapophyseal joints Electronic algometer (kpa)
(Somedic AB, Sweden)
none C5-C6 zygapophyseal joints: 175.4–185.5
Fernández-de-Las-Peñas C et al. 2008 Healthy volunteer C5-C6 zygapophyseal joints Electronic algometer (kpa)
(Somedic AB, Sweden)
none C5-C6 zygapophyseal joints: 308.4–334.5
Ge HY et al. 2006 Healthy volunteer Trapezius
Posterolateral neck
Pressure algometer (kpa)
(Somedic Algometer type 2, Sollentuna, Sweden)
none Trapezius: 320–430
Posterolateral neck: 420–445
Ge HY et al. 2005 Healthy volunteer Trapezius
Posterolateral neck
Pressure algometer (kpa)
(Somedicw Algometer type 2, Sollentuna, Sweden)
none Trapezius: 440–550
Posterolateral neck: 365–405
Nie H et al. 2005 Healthy volunteer Cervical muscle: processus transversus C5
Cervical myotendinous spot: processus transversus C7
Upper trapezius: middle point of processus spinosus C7 and acromion
Levator scapulae: 2 cm superior to the angulus superior scapulae
Angulus superior scapulae
1 cm medial to the acromioclavicular joint
Supraspinatus: 3 cm superior to the middle of spina scapulae
Infraspinatus: 3 cm distal to the middle of spina scapulae
Middle trapezius: middle point of processus spinosus and medial border of spina scapulae
Lower trapezius
Electronic pressure algometer (kpa)
(Somedic Algometer type 2, Sweden)
none Neck/shoulder: Average 322