TABLE 2.

Q Exactive Plus (Thermo Scientific) instrument parameters

Parameter Value
Runtime 0 to 7.5 min
Polarity Positive
Default charge state 1
Full MS
 Resolution 70,000
 AGC targeta 3e6
 Maximum IT 246 ms
 Scan range 100 to 1500 m/z
dd-MS2/dd-SIM
 Resolution 17,500
 AGC target 1e5
 Maximum IT 54 ms
 Loop count 5
 TopN 5
 Isolation window 1.0 m/z
 (N)CE/stepped (N)CE NCE: 20, 40, 60
dd Settings
 Minimum AGC target 8.00e3
 Intensity threshold 1.5e5
 Peptide match Preferred
 Exclude isotopes On
 Dynamic exclusion 10.0s
Tune data
 Spray voltage (+) 3800.00
 Capillary temp (+ or +–) 320.00
 Sheath gas (+ or +–) 35.00
 Aux gas (+ or +–) 10.00
 Spare gas (+ or +–) 0.00
 Max spray current (+) 100.00
 Probe heater temp (+ or +–) 0.00
 S-lens RF level 50.00
 Ion source HESI
a

AGC, Automatic Gain Control; IT, injection time; dd-MS2/dd-SIM, data-dependent MS2/data-dependent selected ion monitoring; NCE, normalized collision energy; RF, radiofrequency.