Skip to main content
. 2013 Mar 1;17(2):247–250. doi: 10.1016/j.bjid.2012.09.014

Table 1.

Results of 1928 specimens detected by two rapid diagnostic kits.

Evaluation kits (Wondfo) Reference kits (GENESIS)
Flu A (n = 1928)
Flu B (n = 1928)
Positive Negative Positive Negative
Positive 246 125a 79 2c
Negative 11b 1546 7d 1840
a

97 cases were confirmed positive.

b

All 11 cases were confirmed negative.

c

1 case was confirmed positive.

d

4 cases were confirmed negative.