Skip to main content
. 2022 Aug 15;9:872705. doi: 10.3389/fsurg.2022.872705

Table 5.

Instructions For Use (IFUs) of last-generation endografts capable of dealing with hostile aortic necks (16, 20, 21, 27, 29, 33, 37).

Device Proximal aortic landing zone
Distal iliac landing zone
Access (minimum vessel diameter)*
Infrarenal landing neck length Aortic neck diameter Aortic neck angulation Iliac neck length Iliac neck diameter Main body Iliac limbs
Alto ≥7 mm 16–30 mm ≤60° ≥10 mm 8–25 mm 5 mm (15 Fr) 4 mm (12 Fr)
Anaconda ≥15mm 16–31 mm
17.5–31 mm (ONE-LOK)
≤90° ≥20 mm 8.5–21 mm 6.667 mm (20 Fr) 6 mm (18 Fr)
Aorfix ≥15 mm 19–29 mm ≤90° ≥15 mm 9–19 mm 6 mm (18 Fr) 5.333 mm (16 Fr)
Conformable ≥15 mm
≥10 mm when aortic neck angulation ≤60˚
16–32 mm ≤90° ≥10 mm 8–25 mm 5 mm (15 Fr) 4 mm (12 Fr)
Endurant II/IIs ≥10 mm or
≥4 mm and <10 mm when used in conjunction with the Heli-FX EndoAnchor
system
19–32 mm ≤60° ≥15 mm 8–25 mm 6 mm (18 Fr)
(5.333 mm (16 Fr) for EndoAnchors)
4.667 mm (14 Fr)
Ovation iX ≥13 mm 16–30 mm ≤45° if proximal neck length <10 mm
≤60° if proximal neck length ≥10mm
≥10 mm 8–25 mm 4.667 mm (14 Fr) 4.333 mm (13 Fr)
Treovance ≥ 10 mm 17–32 mm ≤ 60° if proximal neck length ≥10 mm
≤75° if proximal neck length ≥15 mm
For length of ≥10 mm an inside diameter of 8 mm – 13 mm
or
For length of ≥15 mm an inside diameter of 13 mm – 20 mm
6 mm (18 Fr) 4.333 mm (13 Fr)