Skip to main content
. 2022 Aug 22;9:923544. doi: 10.3389/fcvm.2022.923544

Table 3.

Different LVAD device characteristics and subsequent different emergency management.

Information/devices Heart MATE 3 / II Heartware Jarvik 2000
Mechanism Centrifugal pump with full magnetic levitation of the rotor / axial continue–flow pump Centrifugal flow pump Axial continue–flow pump
Pulse Normally absent or weak Normally absent or weak It could be present depending on myocardial contraction, preload, and afterload
Target vital signs mBP 70–90 mmHg mBP 75–90 mmHg (preferably use the doppler method to measure BP) mBP 65–75 mmHg
Low–flow advices Heart–shaped red light will appear with a continuous acoustic alarm Triangular yellow light with acoustic alarm Low–flow –> Light alarm Pump arrest –> stop signal with red bell and continuous acoustic alarm
Low–flow treatment Hypovolemia –> fluid administration Right HF –> inotropes, fluids, hyperventilation Evaluate if volume expansion is required
Device flow velocity Impossible to speed up in out–of–hospital environment Impossible to speed up in out–of–hospital environment Normally set on 3 velocity, it could be manually adjusted
Heparin therapy Generally, not required (discuss with LVAD implantation center) To decide whether to use heparin, contact LVAD implantation center Generally, not necessary
Defibrillation Possible Possible (don't disconnect device before delivering current) Possible (don't disconnect device before delivering current)
External or manual pump Not present Not present
With ECM, high risk of device displacement: evaluate on clinical basis.
If ECM has to be performed, evaluate pump function and position first
Not present ECM Possible
External pacing Possible Possible Possible
External cables One cable emerges from abdomen One cable emerges from abdomen One cable emerges from retro–auricular area or abdomen
Battery supply Patient should have already been equipped with a set of black batteries (3 h duration) and gray batteries (14–17 / 8–10 h duration; charge conditions could be checked pressing the button on the battery cover) –>At least ONE cable must always be connected to a power generator: DON'T remove simultaneously the two batteries, otherwise the pump will stop Device receive charge from one battery at a time: maximum duration 4–6 h (Both battery and controller have a light signal indicating charge status) Only battery power source (not electrical current) 2 types of battery: Small and portable, 8–10 h duration (could be quantified pressing on the black button on the battery) Big supply battery, minimum 24 h duration

ECM, external cardiac massage; HF, heart failure; LVAD, left ventricular assist device; mBP, mean blood pressure.