Table 3.
Different LVAD device characteristics and subsequent different emergency management.
| Information/devices | Heart MATE 3 / II | Heartware | Jarvik 2000 |
|---|---|---|---|
| Mechanism | Centrifugal pump with full magnetic levitation of the rotor / axial continue–flow pump | Centrifugal flow pump | Axial continue–flow pump |
| Pulse | Normally absent or weak | Normally absent or weak | It could be present depending on myocardial contraction, preload, and afterload |
| Target vital signs | mBP 70–90 mmHg | mBP 75–90 mmHg (preferably use the doppler method to measure BP) | mBP 65–75 mmHg |
| Low–flow advices | Heart–shaped red light will appear with a continuous acoustic alarm | Triangular yellow light with acoustic alarm | Low–flow –> Light alarm Pump arrest –> stop signal with red bell and continuous acoustic alarm |
| Low–flow treatment | Hypovolemia –> fluid administration Right HF –> inotropes, fluids, hyperventilation | Evaluate if volume expansion is required | |
| Device flow velocity | Impossible to speed up in out–of–hospital environment | Impossible to speed up in out–of–hospital environment | Normally set on 3 velocity, it could be manually adjusted |
| Heparin therapy | Generally, not required (discuss with LVAD implantation center) | To decide whether to use heparin, contact LVAD implantation center | Generally, not necessary |
| Defibrillation | Possible | Possible (don't disconnect device before delivering current) | Possible (don't disconnect device before delivering current) |
| External or manual pump | Not present | Not present With ECM, high risk of device displacement: evaluate on clinical basis. If ECM has to be performed, evaluate pump function and position first |
Not present ECM Possible |
| External pacing | Possible | Possible | Possible |
| External cables | One cable emerges from abdomen | One cable emerges from abdomen | One cable emerges from retro–auricular area or abdomen |
| Battery supply | Patient should have already been equipped with a set of black batteries (3 h duration) and gray batteries (14–17 / 8–10 h duration; charge conditions could be checked pressing the button on the battery cover) –>At least ONE cable must always be connected to a power generator: DON'T remove simultaneously the two batteries, otherwise the pump will stop | Device receive charge from one battery at a time: maximum duration 4–6 h (Both battery and controller have a light signal indicating charge status) | Only battery power source (not electrical current) 2 types of battery: Small and portable, 8–10 h duration (could be quantified pressing on the black button on the battery) Big supply battery, minimum 24 h duration |
ECM, external cardiac massage; HF, heart failure; LVAD, left ventricular assist device; mBP, mean blood pressure.