Skip to main content
. 2015 Oct 22;77(5):622–627. doi: 10.1590/S1808-86942011000500015

Frame 2.

ASSR parameters in this study.

Carrying frequency Tone pip at 2 kHz
Stimulus modulation 97 Hz right ear / 93 Hz left ear
Transducer ER 3A in-ear phones
Intensity 20 dBSPL with 10 dBSPL increments to seek the minimum response level
Filters 30 - 300 Hz
Amplification 100,000 times
Sweeps 400 (responses updated every 40)
Detection of responses F statistical technique