Skip to main content
. 2022 Sep 6;77:100105. doi: 10.1016/j.clinsp.2022.100105

Table 1.

Serological indexes of the assays in the present study's samples different systems, using commercial assays defined samples.

Influenza
SARS-CoV 2
IgG IgA IgG IgA
Positive samples 34 34 28 28
Negative samples 29 29 28 28
Sensitivity (95% CI) 0.967 (0.828 to 0.999) 1 (0.880 to 1) 1 (0.868 to 1) 0.966 (0.822 to 0.999)
Specificity (95% CI) 1 (0.894 to 1) 1 (0.897 to 1) 0.933 (0.779 to 0.992) 1 (0.872 to 1)
Positive predictive value (95% CI) 1 (0.881 to 1) 1 (0.881 to 1 0.929 (0.765 to 0.991) 1 (0.877 to 1)
Negative predictive value (95% CI) 0.971 (0.847 to 0.999) 1 (0.897 to 1) 1 (0.877 to 1) 0.964 (0.816 to 0.999)