Table 1.

Detailed design specifications for the PeriPath device

Category Requirement Value Rationale
Size Diameter <1 cm The 1 cm window is essential to deliver larger cardiac therapies to the heart
Mechanical Angles 25 deg Optimal surgical field of view prevents risk of ventricular punctures
Forces Insertion <1 N Device insertion should not require excess force, demonstrating its ease of use
Forces Deployment <10 N Device deployment should not require excess force, demonstrating its ease of use
Forces Working channel drag forces <2 N Manipulation of instrumentation within the working channel should require minimal effort
Forces Visualization channel drag forces >2 N and <10 N Endoscope insertion should not require excess force and should not dislodge during a procedure
Forces Retention >10 N The device should not be easily dislodged when locked into the incision during a procedure
Forces Torque <0.01 N·m The device should not fracture when twisted at a usable torque.
Operation Mode of operation Single User One user should have the ability to hold the endoscope and manipulate instrumentation simultaneously