Size |
Diameter |
<1 cm |
The 1 cm window is essential to deliver larger cardiac therapies to the heart |
Mechanical |
Angles |
25 deg |
Optimal surgical field of view prevents risk of ventricular punctures |
Forces |
Insertion |
<1 N |
Device insertion should not require excess force, demonstrating its ease of use |
Forces |
Deployment |
<10 N |
Device deployment should not require excess force, demonstrating its ease of use |
Forces |
Working channel drag forces |
<2 N |
Manipulation of instrumentation within the working channel should require minimal effort |
Forces |
Visualization channel drag forces |
>2 N and <10 N |
Endoscope insertion should not require excess force and should not dislodge during a procedure |
Forces |
Retention |
>10 N |
The device should not be easily dislodged when locked into the incision during a procedure |
Forces |
Torque |
<0.01 N·m |
The device should not fracture when twisted at a usable torque. |
Operation |
Mode of operation |
Single User |
One user should have the ability to hold the endoscope and manipulate instrumentation simultaneously |