Skip to main content
. 2022 Sep 6;21:63. doi: 10.1186/s12938-022-01032-4

Table 2.

Agreement of in-person auscultation results between the Stemoscope and an acoustic stethoscope among three cardiologists

Cardiologist 1 Cardiologist 2 Cardiologist 3
Acoustic Acoustic Acoustic
Abnormal Normal Abnormal Normal Abnormal Normal
Stemoscope
Abnormal 150 9 148 11 144 15
Normal 9 90 15 84 13 86
Agreement 93% 90% 89%
P value 1 0.44 0.57
Kappa value 0.85 0.78 0.78