Table 6.

Mechanical properties in blood vessels: PCL, TPU, and PLGA tested in vitro (PCL in vivo) and great saphenous vein (GSV).

Test Performed PCL In Vitro TPU In Vitro PLGA In Vitro GSV Reference
Internal Diameter (mm) 4.59 3.83 4 3 [22,23,113,115,117,120,122,123,124,125,132,133,139,140,141,142,145,154,157]
Wall Thickness (µM) 580 540 330 518 [24,25,113,115,117,118,120,123,124,133,140,141,143,145,147,151]
Circumferential Tensile Strength (KPa) 13,360 2570 8350 2405 [12,19,26,113,115,120,122,125,141,143,145,151]
Longitudinal Tensile Strength (KPa) 15,170 11,320 5010 9760 [12,26,115,117,118,122,124,127,132,135,137,139,140,141,142,143,145,147,151,154]
Burst Pressure (KPa) 264.51 451.29 312.84 371.96 [12,26,118,120,122,123,142,143,147,151,154,157]
Suture Retention Strength (g) 1034 803 1050 327 [22,27,115,117,118,123,143,151]
Dynamic Compliance (%/100 mmHg) 5.42 NR 3.41 4.40 [12,28,117,120,124]