Skip to main content
. 2022 Sep 8;17:154. doi: 10.1186/s13014-022-02123-1

Table 2.

Gamma pass rates and mean Gamma values for the (3%/3 mm), (3%/2 mm) and (2%/2 mm) criteria for each tested plan

Treatment modality Gamma criteria Gamma passing rate Mean gamma value

3-Fiducial static

(No target rotation)

(FTTS)

3%/3 mm 99.8% 0.22
3%/2 mm 99.3% 0.29
2%/2 mm 98.6% 0.33

3-Fiducial synchrony

(No target rotation)

(FTTS)

3%/3 mm 96.7% 0.30
3%/2 mm 93.7% 0.39
2%/2 mm 92.9% 0.46

1-Fiducial synchrony

(No target rotation)

(FTTS)

3%/3 mm 99.2% 0.23
3%/2 mm 97.8% 0.30
2%/2 mm 96.9% 0.34

3-Fiducial static

(Target rotation allowed)

(FTTS)

3%/3 mm 98.6% 0.25
3%/2 mm 97.1% 0.33
2%/2 mm 96.3% 0.38

1-Fiducial synchrony

(Target rotation allowed)

(FTTS)

3%/3 mm 96.3% 0.36
3%/2 mm 91.2% 0.48
2%/2 mm 87.8% 0.57

XLTS 2-view static

(Target rotation allowed)

3%/3 mm 99.9% 0.22
3%/2 mm 99.4% 0.28
2%/2 mm 98.4% 0.33

XLTS 2-view synchrony

(Target Rotation Allowed)

3%/3 mm 97.3% 0.35
3%/2 mm 92.8% 0.46
2%/2 mm 89.2% 0.54

XLTS 1-view synchrony

(Target ROTATION ALLOWED)

3%/3 mm 98.5% 0.29
3%/2 mm 96.2% 0.38
2%/2 mm 93.44% 0.44

XLTS 0-View

(Target rotation allowed)

3%/3 mm 85.0% 0.55
3%/2 mm 76.1% 0.70
2%/2 mm 69.7% 0.88