Table 3.

Procedural, electrical, and echocardiographic outcomes

Conduction system pacing (n = 110) Conventional pacing (n = 113) P value
Procedure duration (min) 130 ± 67 101 ± 65 <.01
Fluoroscopy duration (min) 17 ± 12 16 ± 15 .39
Device type
 Pacemaker 91 (83%) 65 (58%) <.01
 ICD 19 (17%) 48 (42%) <.01
 Single-chamber 9 (8%) 21 (19%) <.01
 Dual-chamber 62 (56%) 34 (30%) <.01
 Biventricular 39 (36%) 58 (51%) <.01
Electrical characteristics HBP LBBAP RVP BVP (CS)
 Ventricular pacing leads, n (%) 84 (76%) 46 (42%) 56 (50%) 57 (50%)
 Baseline QRSd (ms) 100 ± 23 111 ± 29 100 ± 21 138 ± 28 <.01
 Paced QRSd (ms) 122 ± 21 128 ± 14 162 ± 21 163 ± 26 <.01
 Pacing threshold (implant) 1.11 ± 0.7 0.63 ± 0.29 0.7 ± 0.33 1.38 ± 1.34
 Pacing threshold (last f/u) 1.1 ± 0.6 0.73 ± 0.22 0.55 ± 0.43 1.2 ± 0.74
 Threshold increase >1 V 6 0 0 3
 Lead revision/abandoned 2 0 5 1
LVEF CSP (n = 77) CP (n = 83)
 Baseline LVEF 46.5 ± 14.2 36.4 ± 16.1 <.01
 Follow-up LVEF 51.9 ± 11.2 39.5 ± 16.0 <.01
Subgroups HBP (58) LBBAP (19) RVP (41) BVP (42)
 Baseline LVEF 46.4 ± 14.1 46.9 ± 15.1 50.3 ± 12.0 26.7 ± 10.5
 Follow-up LVEF 52.3 ± 10.9 50.1 ± 12.9 47.7 ± 13.2 33.8 ± 15.5

CS = coronary sinus; f/u = follow-up; ICD = implantable cardioverter-defibrillator; other abbreviations as in Tables 1 and 2.

P < .05 compared to baseline.