. 2022 Sep 11;8(9):e10591. doi: 10.1016/j.heliyon.2022.e10591

Table 5.

Assay precision.

Target Level (x LoD) Hit rate Mean Ct Instrument to-Instrument
Lot-to-Lot
Day-to-Day
Run-to-Run
Within-Run (Residual)
Total
SD CV% SD CV% SD CV% SD CV% SD CV% SD CV%
Target 1 ∼0.3x 10.0% 32.5 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.5 1.4 0.5 1.4
∼1.0x 91.1% 32.1 0.0 0.0 0.2 0.6 0.1 0.3 0.0 0.0 0.6 1.8 0.6 1.9
∼3.0x 100.0% 31.2 0.0 0.0 0.2 0.7 0.0 0.0 0.0 0.0 0.3 0.9 0.4 1.1
Target 2 ∼0.3x 34.4% 35.4 0.0 0.0 0.5 1.3 0.3 0.8 0.1 0.2 0.5 1.5 0.8 2.2
∼1.0x 93.3% 34.2 0.0 0.0 0.1 0.3 0.2 0.6 0.0 0.0 0.7 2.0 0.7 2.2
∼3.0x 100.0% 32.9 0.0 0.1 0.0 0.0 0.0 0.0 0.0 0.0 0.4 1.1 0.4 1.1